Table 1: The measured and calculated (Anisotropic analytical algorithm and Pencil beam convolution) central axis percent depth doses at P1, P2, and P3 beyond high density heterogeneity interface for field sizes 5 × 5, 10 × 10, and 20 × 20 cm2 (top to bottom) are compared. P1, P2, and P3 are points of interest at depths of 21, 22, and 23 cm, respectively in phantom B [Figure 1]. The measured as well as calculated doses were normalized to the dose of maximum (dmax) reading obtained at 1.5 cm depth. (6 MV photon beam, 100 cm source to surface distance to the surface of phantom)

Table 1: The measured and calculated (Anisotropic analytical algorithm and Pencil beam convolution) central axis percent depth doses at P1, P2, and P3 beyond high density heterogeneity interface for field sizes 5 × 5, 10 × 10, and 20 × 20 cm<sup>2</sup> (top to bottom) are compared. P1, P2, and P3 are points of interest at depths of 21, 22, and 23 cm, respectively in phantom B [Figure 1]. The measured as well as calculated doses were normalized to the dose of maximum (dmax) reading obtained at 1.5 cm depth. (6 MV photon beam, 100 cm source to surface distance to the surface of phantom)