Dose prediction accuracy of anisotropic analytical algorithm and pencil beam convolution algorithm beyond high density heterogeneity interface


Department of Radiation Oncology, Arizona Center for Cancer Care, Peoria, AZ, USA

Correspondence Address:
Suresh B Rana
Department of Radiation Oncology, Arizona Center for Cancer Care, Peoria, AZ
USA
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/2278-330X.105888

Rights and Permissions

Purpose: It is well known that photon beam radiation therapy requires dose calculation algorithms. The objective of this study was to measure and assess the ability of pencil beam convolution (PBC) and anisotropic analytical algorithm (AAA) to predict doses beyond high density heterogeneity. Materials and Methods: An inhomogeneous phantom of five layers was created in Eclipse planning system (version 8.6.15). Each layer of phantom was assigned in terms of water (first or top), air (second), water (third), bone (fourth), and water (fifth or bottom) medium. Depth doses in water (bottom medium) were calculated for 100 monitor units (MUs) with 6 Megavoltage (MV) photon beam for different field sizes using AAA and PBC with heterogeneity correction. Combinations of solid water, Poly Vinyl Chloride (PVC), and Styrofoam were then manufactured to mimic phantoms and doses for 100 MUs were acquired with cylindrical ionization chamber at selected depths beyond high density heterogeneity interface. The measured and calculated depth doses were then compared. Results: AAA's values had better agreement with measurements at all measured depths. Dose overestimation by AAA (up to 5.3%) and by PBC (up to 6.7%) was found to be higher in proximity to the high-density heterogeneity interface, and the dose discrepancies were more pronounced for larger field sizes. The errors in dose estimation by AAA and PBC may be due to improper beam modeling of primary beam attenuation or lateral scatter contributions or combination of both in heterogeneous media that include low and high density materials. Conclusions: AAA is more accurate than PBC for dose calculations in treating deep-seated tumor beyond high-density heterogeneity interface.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed5870    
    Printed124    
    Emailed2    
    PDF Downloaded1086    
    Comments [Add]    
    Cited by others 6    

Recommend this journal